5 kV AC, 6 kV DC Hipot + IR Tester DC Hipot Testers DPC series Udey Test Kits
Description
5 kV AC, 6 kV DC Hipot + IR Tester DC Hipot Testers DPC series Udey Test Kits
Hipot Testers DPC series : UDEY TESTER-US-56I is microprocessor controlled AC/DC Hipot with Insulation Resistance Tester. US-561 is designed to perform high voltage withstand and insulation resistance tests on various types of electrical products, systems and components.
Machine Feature
- AC/DC withstand tests and Insulation resistance tests.
• IR test mode: IR can be performed independently or in sequence with Dielectric Withstand test.
• User selectable output voltage frequency of 50 Hz or 60 Hz.
• Setup of trip current and output voltage without the high voltage activated.
• Store and Recall test set-up memories.
• All parameters are set up through simple menu driven program on screen.
• Line load regulation from no load to full load.
• Front ‘Keys’ can be locked providing limited user access so that only authorized personnel with unlock code can modify test parameters.
• Backlit LCD display allows monitoring of 100uA AC and 20uA DC: Useful when test requirements only allow a very low level of acceptable leakage current.
• Advance A/D conversion, Uses ‘Digital frequency synthesis’ technology to achieve perfect AC test power with crest factor of 1.35-1.45
• Electronic settings: ‘Test Voltage’, ‘High / Low limits of current’, ‘Ramp up time’, ‘Dwell time’.
• Backlit LCM display
• Indicators: Flashing ‘HV ON’, ‘PASS’ and ‘FAIL’ indicators.
• ‘Audio’ alarm at ‘FAIL’.
• Maximum output current is 12mA AC and 6mA DC.
• Remote output for control to configure tester through simple PLC relay control.
• Rugged table top cabinet
Machine Technical Specification
Supply Voltage | 115/230 Volts ± 10%AC, 47-63 Hz single Phase. |
AC WITH STAND TEST | |
Output Test Voltage | Nominal : 0-5kV/12mA AC |
Output Frequency | 50 Hz / 60 Hz |
Output Waveform | Sine wave THD : <2% [Crest Factor > 1.3<1.5] |
TRIP : Upper Limit | 0.10 to 12.00mA; Resolution: 0.01mA per step Accuracy: ± 2% + 2 counts [0.01 to 11mA] |
DC WITH STAND TEST | |
Output Test Voltage | Nominal: 0-6kV/5mA DC |
Output Ripple:mV p-p | <5% at 6kV/5mA Under Resistive Load |
TRIP : Upper Limit | 0.02 to 5.00mA; [0=OFF]. Resolution: 0.01mA per step Accuracy: ± 2% + 2 counts [0.01 to 11mA) |
TRIP : Lower Limit | 0.00 to 5.00mA; [0=OFF]. Resolution: 0.01mA per step Accuracy: ± 2% + 2 counts [0 to 4mA) |
Voltage Setting AC/DC | AC : 0.00 to 5.00kV DC : 0.00 to 6.00kV Resolution : 0.01kV per step; Accuracy: ±2% + 5 Volts. Can be adjusted during test by UP/DOWN Keys |
Voltage Regulation | ± 1% + 5 volts [No load to full load]. |
Ramp Time [ Rise Time] DWELL TIME [Test time] | 0.1 to 999.9 seconds, 0.1 second per step 0=0.1 second fixed setting 0.2 to 999.9 seconds, 0.1 second per step 0=Continuous running |
METERING : 4 digits Voltmeter AC/DC | AC: 0.00 to 5.00kV; DC: 0.00-6.00kV Resolution: 0.01kV; Accuracy:± 2% + 1 count. |
Ammeter AC/DC Timer | AC: 0.10 to 12.00mA; DC: 0.02-5.00mA Resolution: 0.01mA; Accuracy: ± 2% + 1 count. Range: 0.0 to 999.9 seconds; Resolution: 0.1 second. Accuracy: ± 0.1% reading + 0.05 seconds. |
INSULATION RESISTANCE TEST Test Voltage : Settable | 0.01 to 1.00kV DC; Resolution: 0.1kV per step Accuracy: ± 2% + 2 volts. |
Test Voltage : Display | 0.00 to 1.00kV DC; Resolution: 0.01kV per step Accuracy: ± 2% + 1 count. |
Resistance : Display | 1 to 1000 Meg ohm Accuracy : ± 5% + 2 count at voltage = 500V DC. ± 7% + 2 count at voltage = 500V DC. |
Test Time Resistance : Upper Limit Resistance : Lower Limit |
0.5 to 999.9 seconds, 0.1 s per step 0 = Continuous; Accuracy: ± 0.1% + 0.05 s. 0 and 1 to 1000 Meg ohm [0=OFF] Resolution : 1 Meg ohm per step Accuracy : ± 5% + 2 count at voltage = 500V DC. : ± 7% + 2 count at voltage = 500V DC. 1 to 1000 Meg ohm; Resolution : 1Meg ohm per step Accuracy : ± 5% + 2 count at voltage = 500 V DC. : ± 7% + 2 count at voltage = 500 V DC. GENERAL |
Remote Control + Signal Failure Alarms Auto Discharge |
9 pin ‘D’ socket provides the following signals: a) Remote Control : Test, Reset, Interlock b) Outputs : Pass, Fail, Test in progress Buzzer, LCD Display, Indicator Maximum 200 ms Maximum capacitive load verses output voltage 0.2 uF : Maximum 1kV 0.1 uF : Maximum 2kV 0.05 uF : Maximum 4kV 0.04 uF : Maximum 5kV 0.015 uF : Maximum 6kV |
Program Memory | 5 memories, 4 test modes selectable [W, I, W-I, I-W] |
Keyboard Lock | Key Lock to avoid unauthorized access of use. |
Display | LCD, dot matrix, back lit |
Construction Power Cord | Sheet Steel Table Top Cabinet, Powder Coated with tilt-up front Feet. Detachable 2 meter long powder cord terminated in 3 pin grounding plug. |
Use Environment | Temperature : 0-40 deg C; R.H. = Maximum 80% Altitude : Max 2000 Meters. |
Dimensions | W x D x H : 28 x 39 x 9 cm approximately. 9 kg approximately |
Standard Accessories
- a) Mains cord 2 meter long – 1 Number
b) HV and RETURN Test Leads – 1 Set
c) D Connector with cable – 1 Number
Explore the comprehensive range of Udey test kits click here